SVHD (/W) is designed for measurements of very high dose rates (gamma) up to 1000 Sv/h in H*(10), specially designed for the extreme dose rates found in nuclear power plant pools and decommissioning measurements of thin pipes. Withstands accumulated doses of 5 kSv. Minimalist detector head made of stainless steel and equipped with 50 meters of cable. SVHD is available in two variants;

  • SVHD/W for measurements of spent fuel in pools. IP68 and waterproof to 8 atm in borated water.
  • SVHD for decommissioning measurements. IP64 and detector head of only 11 mm.

Data sheet

Contact

Stefan Mårtensson

Stefan

Technical specification

Detector: Energy compensated silicon diode
Measured unit: H*(10)
Energy range: gamma 50 keV – 2 MeV
Measurement range: 1 mSv/h to 1000 Sv/h
Dimension: 11 x 101 mm (Ø x h), samt 17 x 117 mm (Ø x h)
Weight: 100 g
Cable reel: 50 m
Temperature: -30°C to +50°C
Ingress protection: Probe IP68 (SVHD/W), or IP64 (SVHD)
Standards: IEC 60846

 