Föregående
Se fler

LB 9140/LB 9150 – movable alpha/beta particulate monitor moving filter

Tarvitsetko apua?

Autamme sinua löytämään oikean ratkaisun. Tarjoamme tukea, huoltoa, kalibrointia ja koulutusta kaikista tuotteistamme. Ota yhteyttä palvelusopimusta varten!

 

Huolto & Korjaukset Tarjouspyyntö

LB 9140/LB 9150 is a transportable measurement system for airborne alpha and beta emitters in air with compensation for radon and naturally occurring varying background. It can handle air flows up to 3.3 m³/h.

LB 9140 radon compensates with the pseudoincidence method and uses glass fiber filters.
LB 9150 compensation is done with spectrometry which gives lower detection limits. A more stable method when radon varies over time. Membrane filter

The measurement system is built in modules;

Dust collection unit 9140µ
Silicon detector 600 mm²
Cassette with filter: Aluminum IP45. Door made of plexiglass.
Pump – Nominal flow is 3 m³/h (Lp 60 dB(A) at 1m). Flow meter is optional
Data logger LB 5340

Tietolomake

Ota yhteyttä

Toni Kansanoja

Tekniset tiedot

Detector: Si-detector, 600 mm²
Energy range: Alpha from 2,5 MeV, Beta from 100 keV
Detector shielding: 20 mm
Background betachannel from gamma: < 0.4 cps per µSv/h Cs-137 radiation front, < 0.2 cps per µSv/h radiation back
Power: 230V
Datalogger: LB 5340
Flow rate: 3,3 m³/h
Radon compensation: Yes
Filter: 300 x 300 x 530 mm³ (W x D x H)
Filter deposition area: 25 x 25 mm
Filter type LB 9140: Glas fibre Particulates, N° 10, roll 12 m x 50 mm, core Ø25 mm
Filter type LB 9150: Fluoropore membrane filter 40 x 20 mm
Filter speed: selectable 5/10/12.5/15 mm/h
Filter life span: > 3 months at 5 mm/h speed
Temp: -20 to +40°C
Weight: 43 kg


Liittyvät tuotteet

We use cookies on your website to give you the best experience. Read more here.