B ionix – portable tritium monitor
Need help?
We help you find the right solution. We offer support, service, calibration and training for all our products. Contact us for a service agreement!
Service & reparation Request a quote- Tritium detection from 12.5 kBq/m³
- Response time below 75 sec.
- Light and rugged
- Easy to use
- Sounds and lights alarms reporting
- Data storage on an extractable Compact Flash
- Autonomy 6 hours
- Modular
- Manual or automatic gamma compensation
The portable monitor, β ionix is intended for the real-time Tritium activity monitoring and other Beta emitters in ambient air. Due to its very high sensitivity, its ergonomics and its reliability, the β ionix portable monitor provides the radiation levels to your teams on dismantling, construction, process controls, surveillance of premises etc. Ready for use, the portable monitor offers the most advanced features, such as: graphical presentation, archiving of data, the postponement of the alarms, etc
Data sheet
Contact
Göran Jonsson
- Phone: +46 (0)18-56 68 20
- Mobile: +46 (0)70-622 04 07
- goran.jonsson@gammadata.se
Technical specification
Performance (tritium) |
B ionix 3 -MES |
B ionix 3-CMP |
|
Portable tritium monitor with manual gamma compensation |
Portable tritium monitor with automatic gamma compensation |
Measurement range |
2.5 kBq/m³ to 2.5 TBq/m³ |
7 kBq/m³ to 7 TBq/m³ |
Limit of detection |
12.5 kBq/m³ |
25 kBq/m3 |
Precision |
5 % of the reading ± 15 kBq/m³ 25 kBq/m³ / year |
5 % of the reading ± 35 kBq/m³ 25 kBq/m³ / year |
Stability |
25 kBq/m³ / year |
50 kBq/m³ / year |
Zero offset |
± 15 kBq/m³ |
± 35 kBq/m³ |
Noise |
± 12.5 kBq/m³ |
± 25 kBq/m³ |
Response time |
< 60 sec à 90 % of full scale 0.3 %/°C of the reading from 8 to 38 °C |
< 75 sec à 90 % of full scale 0.3 %/°C of the reading from 8 to 38 °C |
Influence of atmospheric pressure |
0.1 %/mBar, hence ± 5 % of the reading from 930 to 1030 mbar |
0.1 %/mBar, hence ± 5 % of the reading from 930 to 1030 mbar |
Influence of humidity |
± 1 % of the reading from 10 to 90 % relative humidity |
± 1 % of the reading from 10 to 90 % relative humidity |
Operation Temperature |
|
|
General characteristics |
|
|
Humidity |
|
|
Volume |
|
|
Ionization chambers |
|
|
Nominal flow rate |
|
|
Std coefficient of response |
|
|
Ionization voltage |
|
|
Related products
M Ionix - mobile tritium monitor
Tritium detection from 10 kBq/m³ Response time below 3 min. HEPA Filtration Easy to use, ready to connect Rugged and Ergonomic Online measurement...
DTionix - on-line tritium system
Tritium detection from 5 kBq/m³ Response time below 15 sec. Ionization chambers from 100cc to 8 L intuitive user interface graphic...