Föregående
Se fler

LB 6360-H10 – Low Dose Rate Proportional Counter

Need help?

We help you find the right solution. We offer support, service, calibration and training for all our products. Contact us for a service agreement!

Service & reparation Request a quote
  • Energy compensated proportional counter tube
  • High voltage for the proportional counter tube (high voltage setting internally or via external control voltage)

The LB 6360-H10 is a proportional counter tube for low dose rate measurements.The measured quantity is ambient dose rate equivalent H*(10). The operating voltage of the counter tube can be set via a control voltage input in order to control the counter tube function (e.g. for plateau recording).

Accessories: Wall mounting bracket (part 25668)

Data sheet

Contact

Stefan Mårtensson

Stefan

Technical specification

Dose rate range: 30 nSv/h – 20 mSv/h
Energy range: 35 keV – 1.3 MeV ± 40 %, with regard to Cs-137 and 0° angular response ± 45°
Intrinsic background approx. 0,12 cps
Calibration factor 0,123 µSv/h per cps
Operating voltage 1700 V
Ambient temperature -20 °C to +60 °C 
Relative humidity 30 to 90 % condensing
Storage temperature max. 60 °C
Outer diameter: 53 mm
Lenght: 550 mm
Weight: 1,6 kg
Protection class: IP65


Related products

We use cookies on your website to give you the best experience. Read more here.