Föregående
Se fler

LB 6701 – ionchamber

Need help?

We help you find the right solution. We offer support, service, calibration and training for all our products. Contact us for a service agreement!

Service & reparation Request a quote

The LB 6701-H10 is an Ionization chamber used for dose rate measurements in various radiation protection applications. The measured quantity are ambient dose and rate equivalent H*(10). The ionization chamber is made from Aluminium, with a Nitrogen gas filling and a radiation resistance up to 106 Gy. Current measuring, converting to pulses.

The ionization chamber exists in 2 versions:

  • LB 6701L-H10 Low dose rate 10 µSv/h to 10 Sv/h
  • LB 6701H-H10 High dose rate 100 µSv/h to 100 Sv/h

A Sr-90 with 50 kBq check source is built in the Ionization Chamber to continuously monitor the proper functioning of the system.

Contact

Stefan Mårtensson

Stefan

Technical specification

Measuring range:
LB 6701L-H10 = 10 µSv/h – 10 Sv/h
LB 6701M-H10 = 0,1 mSv/h – 100 Sv/h
Energy range: 45  – 1300 keV
Calibration factor: 1,4 fA/µSv/h
High Voltage 1000 V
Ambient temperature 0°C to +50°C
Relative humidity 20 to 80 % condensing
Dimension Ionisation Chamber: Ø80 x 268 mm
I-F-Converter Connection box: 240 mm x 160 mm x 90 mm
Protection class: IP 65


Related products